The right accessories let one pendulum test to more than one standard. Strikers, anvils, and specimen supports set the geometry the standard requires. Cooling baths and tongs keep low-temperature specimens at temperature until impact. Gauges and hardware keep the machine aligned and anchored. All accessories on this page fit the Instron® MPX Series Pendulum Impact Testing System. Most also fit the SATEC™ SI Series pendulums still in service.
Charpy Testing Accessories
The striker and anvils set the geometry of a Charpy test, and the geometry changes the absorbed energy. ASTM E23 specifies an 8 mm striker radius. ISO 148-1 allows a 2 mm or 8 mm radius, and results from the two are not interchangeable. Choose the kit for the standard your customer or specification requires.
Each kit includes a striker, left and right anvil inserts with bolts, and shrouds. Use a kit to set up a new standard, or to replace worn parts together.
Source: Accessories Catalog, page 311
Izod Testing Accessories
An Izod test clamps the specimen upright and strikes the free end, above the notch. Some material specifications and customer requirements call for Izod results instead of Charpy results. With the conversion kit and an Izod striker, the same MPX Series frame runs Izod tests to ASTM E23 as well as Charpy tests, so you do not need a second machine or the floor space for one.
Low-Temperature Testing Accessories
Steels change from ductile to brittle as temperature drops, so many specifications call for Charpy tests below room temperature. Results depend on the specimen being at the set temperature at the moment of impact. ASTM E23 and ISO 148-1 require the specimen to be broken within 5 seconds of leaving the cooling medium. The MPX Series starts the test automatically when the door closes, which helps operators stay within that limit.
The bath holds a rack of up to 65 specimens, so a full set conditions at once. Its temperature stability meets the requirements of ASTM E23 and ISO 148-1.
Performance figures are based on methanol at an ambient room temperature of about 22°C (72°F). A warmer room or a different fluid may raise the lowest temperature the bath can reach.
Specimen Preparation and Measurement
Notch geometry and post-test measurements affect results as much as the machine does. A notch that is too shallow, too blunt, or off-center changes how the specimen breaks, and two labs testing the same steel can report different energies. Measuring the broken specimen adds information that absorbed energy alone does not give, such as how much the material deformed before fracture. These tools work with any Charpy impact tester, so you can use them alongside an MPX Series system or an older machine.
Source: Accessories Catalog, page 314
Installation and Maintenance
A pendulum must be anchored to a rigid foundation. If the frame moves during impact, some energy goes into the floor instead of the specimen. Use the template to place the anchor bolts before the concrete is poured.
Source: Accessories Catalog, page 314
SI Series Note
Most strikers, anvils, specimen supports, tongs, and gauges on this page also fit SATEC SI Series pendulums. For SI Series support status, see Out of Production Impact Drop Towers and Pendulums.